DATA ENTRY CONTROLS MENU (CONTINUED)
Hydrogen Entry Mode:
This key steps through the options for entering hydro-
gen content for calculating the net heat of combustion
either manually through the touch screen or automati-
cally through the balance (USB) port.
Oxygen Entry Mode:
This key steps through the options for entering oxygen
content for calculating the net heat of combustion either
manually through the touch screen or automatically
through the balance (USB) port.
Nitrogen Entry Mode:
This key steps through the options for entering nitrogen
content for calculating the net heat of combustion either
manually through the touch screen or automatically
through the balance (USB) port.
Auto Sample ID Controls:
Accesses sub-menu for controlling the automatic assign-
ment of sample identification numbers.
Automatic Sample ID. When set to ON the unit
will automatically assign sample identification
numbers in accordance with parameters set by
the other three keys on this menu. When set to
OFF, the user manually enters each sample ID
when prompted to do so.
Auto Sample ID Prefix. An entry here will be used
as a prefix for all sample IDs, if the Automatic
Sample ID is set to ON. Press this key to access
a sub-menu for entering an alphanumeric prefix.
Next Auto Sample ID Number. Establishes the
initial sample number for a series of tests and
then shows the next sample ID which will be as-
signed. Used when the Automatic Sample ID is
set to ON. Press this key to access a sub-menu
for entering a numeric increment.
Auto Sample ID Increment. Establishes the incre-
ment between sample numbers; used when the
Automatic Sample ID is set to ON. Press this
key to access a sub-menu for entering a numeric
increment.
Sample Weight Warning Above:
This key displays and leads to a sub-menu used to set
the maximum allowable sample weight (including
spike) in grams. A warning will be given if sample
weights above this value are entered.
Spike Weight Entry Mode:
This key steps through the options for entering spike
weights either manually through the touch screen, bal-
ance (USB) port or network.
Sulfur Entry Mode:
This key steps through the options for entering the sul-
fur correction value either manually through the touch
screen or automatically through the balance (USB) port.
Moisture as Determined Entry Mode:
This key steps through the options for entering the
moisture as determined correction value either manual-
ly or through the touch screen or automatically through
the balance (USB) port.