6510 Water Handling System Operating Instruction Manual
- 7 -
Preventative Maintenance
All power should be disconnected when cleaning or servicing the instrument. The
following preventative maintenance for the 6510 Water Handling System is
recommended:
Monitor the amount of water in the water tank and add water as required to
maintain the water level in the reservoir.
Clean the grill on the heat sink at the rear of the Water Handler monthly. Using
compressed air to perform this is a quick and easy way to remove the dust. Wipe
exterior surfaces with lightly dampened cloth containing mild soap solution.
Change the water in the tank monthly.
Change the in-line filter, 149C, every 3 months
Replace the
3
/
8
OD tubing, HX0038TB062, annually
Add an anti-microbial solution, such as VWR Clear Bath Algicide (13272-031),
per the manufacturer’s instructions.
The pipet can be cleaned with a 1:10 dilution of Acetic Acid or straight vinegar.
The reservoir can be easily emptied by repeatedly draining the pipet until the circulating
pump is sucking air. When this occurs, turn the power switch at the rear of the system
to the OFF position. Do not let the pump run dry for more than a few minutes.
There are no user serviceable parts inside the product other than what is specifically
called out and discussed in this manual. Advanced troubleshooting instructions beyond
the scope of this manual can be obtained by calling Parr Instrument Company in order
to determine which part(s) may be replaced or serviced.
Fuses
Lines protective fuses rated: Slo-Blo 4.0 Amp, 250VAC (Parr number 139E20)
Fast-Act 15 Amp, 250VAC (Parr number 139E23)
The replacement of protective fuses for the 6510 Water Handling System should be
performed by qualified personnel.