Operating Instruction Manual 592M 6725 Semi-micro Oxygen Bomb Calorimeter Operating Instruction Manual For models produced after November 2010
6725 Semi-micro Calorimeter Instruction Manual TABLE OF CONTENTS TABLE OF CONTENTS ................................................................................................1-1 PREFACE ......................................................................................................................1-1 Scope .........................................................................................................................1-1 Explanation of Symbols..............................................
6725 Semi-micro Calorimeter Instruction Manual File Management Menu ...........................................................................................5-17 Run Data File Manager ............................................................................................5-18 Diagnostics Menu.....................................................................................................5-19 CALCULATIONS...............................................................................................
6725 Semi-micro Calorimeter Instruction Manual PREFACE Scope This manual contains instructions for installing and operating the Parr 6725 Semi-micro Calorimeter. For ease of use, the manual is divided into 13 chapters. Installation Quick Start Operation Menu Descriptions Calculations Reports Memory Management Maintenance Troubleshooting Technical Service Parts Lists Drawings Tables Subsections of these chapters are identified in the Table of Contents.
6725 Semi-micro Calorimeter Instruction Manual Explanation of Symbols I On position O Off position ~ Alternating Current (AC) This CAUTION symbol may be present on the Product Instrumentation and literature. If present on the product, the user must consult the appropriate part of the accompanying product literature for more information. ATTENTION, Electrostatic Discharge (ESD) hazards. Observe precautions for handling electrostatic sensitive devices. Protective Earth (PE) terminal.
725 Semi-micro Calorimeter Instruction Manual Intended Usage If the instrument is used in a manner not specified by Parr Instrument Company, the protection provided by the equipment may be impaired. General Specifications Electrical Ratings 115VAC, 2.0 Amps. 50/60 Hz 230VAC, 2.
6725 Semi-micro Calorimeter Instruction Manual INSTALLATION The 6725 Semi-micro Oxygen Bomb Calorimeter is to be used indoors. It requires at least 4 square feet of workspace on a sturdy bench or table free from room drafts or radiant heat sources, (preferably in an air conditioned room providing minimal temperature change), an electrical outlet, a supply of water which is at least 1 °C cooler then room temperature and a waste drain will be required.
6725 Semi-micro Calorimeter Instruction Manual Figure 1 6772 Calorimetric Thermometer Back Panel 493E Plug with 3280HC Retainer installed to insulate from live circuits when no motor is attached. Oxygen Tank Installation Install the oxygen tank according to the manufacturer’s recommendations.
6725 Semi-micro Calorimeter Instruction Manual instrument. Install the thermistor in the cover opening and press the bushing firmly into place to anchor the probe in its proper position. Place the cover on the calorimeter with the orienting pin in the alignment hole. The 6772 Calorimetric Thermometer acts an ignition unit for automatically firing the bomb in the 6725 calorimeter.
6725 Semi-micro Calorimeter Instruction Manual The format of an unknown balance can be determined by logging the balance output to the printer attached to the Calorimetric Thermometer. Those protocols which send a command string to the balance will do so while logging is active. In order for the logging to produce meaningful results, the cable connecting the balance to the balance input port of the Calorimetric Thermometer must be correctly wired or configured.
6725 Semi-micro Calorimeter Instruction Manual Bar Code Port The use of barcodes in the laboratory has become a highly accurate, rapid and inexpensive way to identify samples. When purchasing this feature, the user must supply Parr with the MAC address of the calorimeter (found in the Software & Hardware Info menu screen). This allows Parr to activate the feature key. In order to enable the calorimeter to use the bar code feature, the feature key needs to be entered into the instrument.
6725 Semi-micro Calorimeter Instruction Manual The following screenshot illustrates the contents of the Calorimetric Thermometer data directory as presented by a web browser. You can drag and drop or copy and paste test data files (with the csv suffix) from the web browser window to any convenient folder or directory on the PC.
6725 Semi-micro Calorimeter Instruction Manual The Calorimetric Thermometer offers a web server service. Test reports can be viewed with a web browser using a URL of the following form. http://192.168.0.125 Where 192.168.0.125 is the IP address of the Calorimetric Thermometer. The following screenshot illustrates the Calorimetric Thermometer status page.
6725 Semi-micro Calorimeter Instruction Manual Clicking on the Config tab displays some of the configuration options available in the 6772. Changes made on this page will change the settings in the 6772.
6725 Semi-micro Calorimeter Instruction Manual Clicking on the Run Data tab displays a list of reports currently in the instrument memory.
6725 Semi-micro Calorimeter Instruction Manual Clicking on System Info will bring up the Software and Hardware Info screen from the 6772.
6725 Semi-micro Calorimeter Instruction Manual Clicking on the LCD Snapshot Page will bring up an image of the current display on the 6772. Note that this is a picture only. To operate the 6772 from a PC a Remote Feature Key must be installed. Please contact Parr Instrument Company for more details.
6725 Semi-micro Calorimeter Instruction Manual Clicking on Documentation will provide links to the manual and other helpful documents. Note: a connection to the Internet is required to access the documentation.
6725 Semi-micro Calorimeter Instruction Manual QUICK START Detailed Instructions for preparing the sample and charging the 1109A Oxygen Bomb are given in Operating Instructions Manual No. 492M. Follow these instructions carefully, giving particular attention to the precautions to be observed in charging and handling the bomb. The default values of the 6772 are designed to operate with the 1341 Plain Jacket calorimeter. Note the parameter changes in the Operation section of this manual. 1.
6725 Semi-micro Calorimeter Instruction Manual 9. At the conclusion of the test, the calorimeter will signal the user. 10. Open the cover and remove the bomb and Dewar flask. Remove the bomb from the flask and release the residual gas pressure before attempting to remove the cap. This release should proceed slowly over a period of not less than one minute to avoid entrainment losses.
6725 Semi-micro Calorimeter Instruction Manual OPERATION Menu System All configurations and operations are handled by a menu-driven system operated from the bright touch screen display. The settings and controls are organized into ten main sections as displayed on the MAIN MENU. Note: Keys with a “double box” in the upper left hand corner lead to sub-menus. Menu Keys The controls that change the data field information in the menus will be one of the following: 1. Toggles.
6725 Semi-micro Calorimeter Instruction Manual 5. Help. This key is used to access help screens related to the menu currently displayed on the touch screen. 6. This key appears in the Escape key location when the main menu is displayed. This key is used to shut down the calorimeter program before turning off the power.
6725 Semi-micro Calorimeter Instruction Manual The default parameters of the 6772 Calorimetric Thermometer can be changed to guarantee that the thermometer, when cold restarted, will always be in the desired configuration before beginning a series of tests. Users who wish to permanently revise their default settings may do so using the following procedure: • • • Establish the operating parameters to be stored as the user default settings.
6725 Semi-micro Calorimeter Instruction Manual Manual Test Sequencing Some users may wish to construct their own thermogram and apply the classic graphical corrections developed by Dickenson and others. In this case the actual temperatures can be logged to the thermometer memory of the thermometer and then analyzed at the end of the test.
6725 Semi-micro Calorimeter Instruction Manual Figure 2 Cross-Section of the 6725 Semi-micro Calorimeter Install the thermistor in the cover opening and press the bushing firmly into place to anchor the probe in its proper position. Place the cover on the calorimeter with the orienting pin in the alignment hole. Turn the stirrer drive by hand to be sure that it runs freely, then slip the drive belt first onto the motor pulley and then onto the stirrer pulley.
6725 Semi-micro Calorimeter Instruction Manual should begin to significantly rise at this point, indicating sample combustion. The calorimetric post-period should last for an additional 6-7 minutes. At the conclusion of the post-period, turn the stirrer off by pressing the stirrer key once again. The motor switch may be left in the “on” position for subsequent tests. If the data log destination is a log file, the log file is located at /flash/datalog.csv and may be retrieved via FTP.
6725 Semi-micro Calorimeter Instruction Manual The calorimeter temperature should begin to significantly rise at this point, indicating sample combustion. The calorimetric post-period will last for an additional 9-10 minutes until the calorimeter temperature drift rate sufficiently stabilizes.
6725 Semi-micro Calorimeter Instruction Manual MENU DESCRIPTIONS Note: Keys which make global changes to the setup of the calorimeter contain a YES or NO response to make certain that the user wishes to proceed. This two step entry is intended to prevent inadvertent global program changes. Main Menu Selecting the Main Menu key on any menu will return you to the screen pictured on the right of this page. : This key appears in the Escape key location when the main menu is displayed.
6725 Semi-micro Calorimeter Instruction Manual Start Preweigh: This key is used to start the sample pre-weigh process. The user is presented with or prompted for a sample ID. Next, the user is asked to key in the associated sample mass or alternatively the mass is retrieved from a connected balance. Stirrer: On / Off This key provides a convenient way to manually start and stop the calorimeter stirrer motor. The motor must be physically turned on for this function to be active.
6725 Semi-micro Calorimeter Instruction Manual Bucket Plot Color: Toggles between: • Red • Green • Yellow • Blue • Magenta • Cyan • White • Black Bucket Max Value: Press this key to access its numeric dialog box to set a maximum bucket value. Jacket Plot Color: Toggles between:(same as Bucket Plot Color, above). Jacket Max Value: Press this key to access its numeric dialog box to set a maximum jacket value.
6725 Semi-micro Calorimeter Instruction Manual Jacket Temp. Source: Press this key to display a menu that allows the user to select the source for the jacket temperature used for tests. “Other” Multiplier: Press this key to display the Other Multiplier dialog box, where the user can enter a final multiplier to be used when the reporting units are set to “Other”. Calibrate Touchscreen: This key prompts the user to touch the screen at predefined points in order to facilitate touch screen calibration.
6725 Semi-micro Calorimeter Instruction Manual Weight of Fixed Spike: The weight of the fixed spike is entered on sub-menu keyboard. Normally the Calorimetric Thermometer will prompt the user for the weight of the sample and then the weight of the spike. Prompt for Spike Before Weight: When set to ON, the Calorimetric Thermometer will prompt the user for the weight of the sample and the weight of the spike.
6725 Semi-micro Calorimeter Instruction Manual Settings Protect: Provides protection for the program options and settings on the menus. If this is turned ON, the user will be warned that enumeration keys are locked when a key is pressed. Enumeration Keys either toggle a value (ON/OFF) or select from a predefined list. This feature is used primarily to protect the instrument settings from accidental changes if one were to inadvertently touch or bump up against the touch screen.
6725 Semi-micro Calorimeter Instruction Manual User Function Setup: This key leads to sub menus that support the configuration of five factory/user definable function keys. The function keys are accessible from the Diagnostics page. Cold Restart: This is essentially the same as cycling power on the unit. All valid test data will be retained during this cold restart procedure.
6725 Semi-micro Calorimeter Instruction Manual • Charted Value: Toggles the charted value between the HOC Standard (Heat Of Combustion of Standard) and Energy Equivalent. • Process Sigma: In relation to calorimetry, sigma is used as the classification of the instrument. The higher the process sigma the higher the limits for acceptable values for precision control. The Process Sigma of the 6772 will depend on the application. • Temp.
6725 Semi-micro Calorimeter Instruction Manual limit set by Bomb Service Interval (on the Calibration Data and Controls screen), the user will be informed that the bomb is ready for service. Name: Enables the operator to assign a unique alpha-numeric label for the bomb ID. The ID can be up to 8 characters. Protect EE Value: Toggles between OFF and ON. When set to OFF, the 6772 automatically updates the EE value as new tests are run.
6725 Semi-micro Calorimeter Instruction Manual Thermochemical Corrections Menu The Thermochemical Corrections Menu permits three types of fixed corrections for standardization (instrument calibration) runs, and the same three types for determination (test) runs. Pressing the LEFT side of each key toggles the correction ON or OFF. Press the RIGHT side of each key to access the specific numeric dialog box where that fixed value can be set.
6725 Semi-micro Calorimeter Instruction Manual Calculation Factors: Accesses the Calculation Factors sub-menu, which provides for setting a number of options for the way the thermochemical corrections are applied. Determination Correction Fixed Fuse Correction: Press this key on the LEFT side to toggle ON or OFF the fixed fuse correction for determination runs.
6725 Semi-micro Calorimeter Instruction Manual Calculation Factors Menu Nitric Acid Factor: Ratio of the nitric acid correction when the released energy corresponds to 6318 calories. The default is 1.58 calories per 1000 calories of released energy. Acid Multiplier: This multiplier is the normality of the sodium carbonate used to titrate for the acid correction. The default value of 0.0709 allows for direct entry of the acid correction in calories.
6725 Semi-micro Calorimeter Instruction Manual Net Heat/Dry Heat Factors Fixed Hydrogen: ON/OFF and value entry. Fixed Oxygen: ON/OFF and value entry. Fixed Nitrogen: ON/OFF and value entry. Calculate Net Heat of Combustion: ON/OFF. Turn On to have the calorimeter calculate the net heat of combustion. Fixed Moisture as Determined: ON/OFF and value entry. Fixed Moisture as Received: ON/OFF and value entry. Dry Calculation: ON/OFF. Turn on to have the calorimeter calculate the result on a dry basis.
6725 Semi-micro Calorimeter Instruction Manual Automatic Sample ID: When set to ON the unit will automatically assign sample identification numbers in accordance with parameters set by the other three keys on this menu. When set to OFF, the user manually enters each sample ID when prompted to do so. Auto Sample ID Prefix: An entry here will be used as a prefix for all sample IDs, if the Automatic Sample ID is set to ON. Press this key to access a sub-menu for entering an alphanumeric prefix.
6725 Semi-micro Calorimeter Instruction Manual Auto Preweigh Controls: Accesses sub-menu, used to automatically assign Sample ID numbers when a series of samples are preweighed ahead of the time they are actually tested. Automatic Preweigh ID: ON/OFF toggle for this feature. Automatic Preweigh ID Prefix: An entry here will be used as a prefix for all pre-weigh sample IDs.
6725 Semi-micro Calorimeter Instruction Manual Recalculate Final Reports: When set to ON, causes a recalculation of stored final reports using calibration data and menu settings currently in the Calorimetric Thermometer. When set to OFF, the stored final reports are not affected by the current run. Use New EE Value in Recalculation: When set to ON, any recalculation made will use the most recent EE value in the calculations.
6725 Semi-micro Calorimeter Instruction Manual • • • • Data Characters from Balance. This setting is only used when the generic balance format is selected. This value determines the number of numeric data characters (0-9 . + -) to accept. Any additional characters after this value and before the string terminating are discarded. Data Precision. This key allows the user to establish the number of digits to the right of the decimal point that are passed from the balance handler.
6725 Semi-micro Calorimeter Instruction Manual Copy Run Data to SD Card: This key copies all test data to a SD memory card inserted into the rear of the calorimeter controller. This feature is used as a means of either archiving data or transferring it to a PC. Copy User Settings to SD Card: This key copies all previously saved user setups to the SD. Copy User Settings From SD Card: This key copies all user setups previously saved to SD back to the calorimeter controller memory.
6725 Semi-micro Calorimeter Instruction Manual Diagnostics Menu Provides the user with the means to test many of the components and subsystems of the Calorimetric Thermometer. These capabilities should be used in conjunction with this instruction manual in order to obtain the maximum benefits from these capabilities. Calorimetry Parameters: This key accesses its sub-menu which allows access to the calorimetry parameters used to automatically sequence a calorimetric test.
6725 Semi-micro Calorimeter Instruction Manual Data Log Destination: Options are logfile, printer or both. When the logfile option is selected, the logfile is located at /flash/datalog.csv. The maximum allowed size for this file is roughly one megabyte. If the file reaches this size, logging is halted. Select Data Log Items: Press this key to access the Data Log Items sub-menu, which provides keys for fifteen items that can be individually selected for logging.
6725 Semi-micro Calorimeter Instruction Manual Instrument Monitor: This key accesses its sub-menu screen which provides a summary of most of the important instrument parameters. This screen is used to detail the course of a test or to observe the heating / cooling performance of the calorimeter.
6725 Semi-micro Calorimeter Instruction Manual CALCULATIONS Corrections Fuse Correction The fuse correction applied by the Calorimetric Thermometer is calculated as: e3 = (fuse value) (Fuse Multiplier) Fuse corrections can be entered when Fixed Fuse in the Thermochemical Corrections menu, is set to OFF. To enter in the fuse correction in cm while using the Parr 45C10 fuse wire, enter a Fuse Multipler of 2.3 cal/cm. The calorimeter will perform the calculation and provide the correction in calories.
6725 Semi-micro Calorimeter Instruction Manual In practice, the amount of heat produced by the combustion of Nitrogen in the 22 mL and 1109A bomb is negligible and is usually handled by using the Fixed HNO3 Correction for Nitrogen and setting it to 1.5. Sulfur Correction In the oxygen rich atmosphere within the bomb, sulfur in the sample is oxidized to sulfur trioxide which combines with water vapor to form sulfuric acid.
6725 Semi-micro Calorimeter Instruction Manual reasonably constant sulfur content. Any value set-up as a fixed correction will be automatically applied and the controller will not prompt the user for this value. Definitions Total acid is the amount of base required to titrate the bomb washings (milliliters). Nitric acid is that portion of the total acid in the bomb washings that result when the nitrogen in the air that is trapped in the bomb is burned at high pressure.
6725 Semi-micro Calorimeter Instruction Manual Example: Where: 16.03 is equal to the equivalent weight of sulfur in H2SO4 e1 is the nitric acid portion of the total acid correction e2 is the sulfuric acid portion of the total acid correction e1+e2 is the total bomb acid correction Sulfur Multiplier Calculation: %Sulfur × Sample mass × 10 = Sulfur (mg) Sulfur (mg) = milliequiv alents of H 2 SO 4 equivalent weight of sulfur in H 2 SO 4 10 = 0.
6725 Semi-micro Calorimeter Instruction Manual Fixed Total: The Acid Correction represents the total base required to titrate the bomb washings (in milliliters). This includes both nitric and sulfuric acid. The correction is a fixed value set by the operator. The calculation is: e1 = [((total acid)(acid multiplier)) – (% sulfur)(sample mass)(sulfur multiplier)](heat of formation of nitric acid) Using the default acid and sulfur multipliers as well as a heat of formation of nitric acid of 14.
6725 Semi-micro Calorimeter Instruction Manual Note: The values entered into the test report appear as entered in the report. Values for e1, e2 and e3 are calculated and used as energy corrections in accordance with the formulas and settings given above. The formulas used above to arrive at e1 or e2 are not the same as the formulas used for e1 and e2 which appear in most ASTM bomb calorimetric procedures. However, the sum of e1 and e2, above, is equal to the sum of the ASTM treatment of e1 and e2.
6725 Semi-micro Calorimeter Instruction Manual ASTM, ISO and Other Methods Current ASTM, ISO, and British Standard Methods differ on their treatment of the nitric and sulfuric acid thermochemical corrections. ASTM Methods call for titrating the bomb washings to determine the total acid present. This is assumed to be all nitric acid with a heat of combustion of -14.1 Kcal per mole. The amount of sulfur is then determined and converted to equivalents of sulfuric acid.
6725 Semi-micro Calorimeter Instruction Manual The values for fixed acid and sulfur, which are used in preliminary reports, will reflect a sulfur correction of 0, and a nitric acid correction of 10 calories. Conversion to Other Moisture Bases The calculations described previously give the calorific value of the sample with moisture as it existed when the sample was weighed. For example, if an air-dried coal sample was tested, the results will be in terms of heat units per weight of air-dry sample.
6725 Semi-micro Calorimeter Instruction Manual REPORTS The 6772 Calorimetric Thermometer can transmit its stored test data in either of two ways. The Auto Report Destination key on the Reporting Controls Menu toggles the report destination between the display and an optional printer connected to the RS232 printer port of the Calorimetric Thermometer. This menu also selects the type of reports that are generated automatically by the Calorimetric Thermometer.
6725 Semi-micro Calorimeter Instruction Manual To bring the selected report or series of reports to the display, press the DISPLAY key. To send the reports to the printer press the PRINT key. The EDIT key brings up a sub-menu which enables the operator to edit any of the data in the report or add thermochemical corrections to convert preliminary reports to final reports. Final reports can only be edited if Edit Final Reports on the Reporting Controls Menu is turned on.
6725 Semi-micro Calorimeter Instruction Manual MEMORY MANAGEMENT The 6772 Calorimetric Thermometer will hold data for 1000 tests in its memory. These tests may be pre weights, preliminary or final reports for either calibration or determination runs. Once the memory of the controller is filled, the controller will not start a new analysis until the user clears some of the memory. The FILE MANAGEMENT key on the main menu leads to the file management sub-menu.
6725 Semi-micro Calorimeter Instruction Manual MAINTENANCE Refer to Manual 492M, Operating Instructions for the 1109A Semi-micro Oxygen Bomb for bomb maintenance information. CAUTION: Risk of Electrical Shock Disconnect the electrical power before servicing or replacing any components! Fuses The replacement of protective fuses for the 6722 Calorimetric Thermometer should be performed by qualified personnel.
6725 Semi-micro Calorimeter Instruction Manual TROUBLESHOOTING Error List The calorimeter will run a number of diagnostic checks upon itself and will advise the operator if it detects any error conditions. Most of these errors and reports will be selfexplanatory. The following list contains errors that are not necessarily self-evident and suggestions for correcting the error condition.
6725 Semi-micro Calorimeter Instruction Manual Probe substitution can be useful in determining the cause of the problem (probe or electronics). The valid working range of the probe resistance is 1000 to 5000 ohms. • • • Check connection to board. Check quick disconnect. Replace probe. A/D Initialization Failed. Shortly after power is applied to the calorimeter controller and the operating system has started, the CPU attempts to read the unique I/O board calibration information from the I/O board.
6725 Semi-micro Calorimeter Instruction Manual TECHNICAL SERVICE Contact Us Should you need assistance in the operation or service of your instrument, please contact the Technical Service Department. Telephone: (309) 762-7716 Toll Free: 1-800-872-7720 Fax: (309) 762-9453 Email: parr@parrinst.com Any correspondence must include the following basic information: 1. 2. 3. 4. The model and serial # of the instrument. Date purchased. Software version(s) shown on the “Software and Hardware Information” page.
6725 Semi-micro Calorimeter Instruction Manual PARTS LISTS Principal Assemblies in 6772 Calorimetric Thermometer Part Number 1940E A2141E A2154E4 A2140E A2166E A2167E 2147E 1168E2 A719E 1200EEE 1889E 1344DD 139E23 1641E Description Power Supply LCD Transition Board CPU Board I/O Board Cable, I/O to CPU USB Cable, USB Peripheral Touchscreen LCD with Cable Thermistor Probe Cordset, 115V Corset, 220V, EU Plug Film, Display Protection Stylus Fuse Fast/ Act 15 Amp 250V Fuse Fast/ Act 1.
6725 Semi-micro Calorimeter Instruction Manual Parts for the 6725 Semi-micro Calorimeter Part Number 1824 A338CEB A338CEE 45C10 118C2 119C2 122C2 123C2 127C2 135C 139C2 23AC 165AC2GE 208AC A468E A719E 1200EEE 752HC2 A67C2 A1195E2 141C A137C A72C2 3414 534A A38A 533A 6009(A) Description Oxygen Filling Connection Stirrer Drive Motor Assembly 115V Stirrer Drive Motor Assembly 230V Fuse Wire, NI Alloy 10CM Bearing Tower Cover Locating Ring (6725) Dewar Flask 894mL Coupling (6725) Drive Belt, Geared O-ring, Bun
6725 Semi-micro Calorimeter Instruction Manual DRAWINGS 6772 Schematic 13-1
25 Semi-micro Calorimeter Instruction Manual Internal Parts View of 6772 Calorimetric Thermometer 13-2
6725 Semi-micro Calorimeter Instruction Manual Internal Parts View of 6772 Calorimetric Thermometer Note: 1262DD Capacitor Bracket and 897E Capacitor not shown for clarity.
6725 Semi-micro Calorimeter Instruction Manual Back Panel of 6772 Calorimetric Thermometer 493E Plug with 3280HC Retainer installed to insulate from live circuits when no motor is attached.
6725 Semi-micro Calorimeter Instruction Manual External Parts View of the 6725 Semi-micro Calorimeter 13-5
6725 Semi-micro Calorimeter Instruction Manual Internal Parts Views of 6725 Semi-micro Calorimeter 13-6
6725 Semi-micro Calorimeter Instruction Manual 1824 Oxygen Filling Connection 13-7
6725 Semi-micro Calorimeter Instruction Manual TABLES Table 1 – Default Settings Note: The following are recommended operation settings for the 6725 Semi-micro Calorimeter. Calorimeter Operation Operating Mode Bomb Installed/EE Stirrer Determination 1/2400 OFF Operating Controls Method of Operation Reporting Units Use Spiking Correction 4.
6725 Semi-micro Calorimeter Instruction Manual Table 1 – Continued Auto Preweigh Controls Automatic Preweigh ID Automatic Preweigh ID increment Automatic Preweigh ID Number Reporting Controls Report Width Automatic Reporting Auto Report Destination Individual Printed Reports Edit Final Reports Recalculate Final Reports Use New EE Values in Recalculation Communication Controls Printer Printer Type Printer Destination Diagnostics Calorimetry Parameters Tolerance (L) Parameters Preperiod Tolerance (L1) Mis-Fi
6725 Semi-micro Calorimeter Instruction Manual Table 2 – ISO & BSI Method Settings Page Line Acid Correction (STD) Thermochemical Corrections Fixed Sulfur STD Acid Correction (DET) Calculations Factors Fixed Sulfur DET Acid Multiplier Sulfur Value is Percent Sulfur Multiplier Use Offset Correction Offset Value 14-3 Setting Entered HNO3 Off Entered HNO3 Off Value 13 7 13 7 0.154 Off 0.1 On -43.
PARR INSTRUMENT COMPANY 211 Fifty - Third Street Moline, Illinois 61265 USA 309 / 762-7716 800 / 872-7720 Fax 309 / 762-9453 http://www.parrinst.com parr@parrinst.