Manual
TDM Installation
Parr Instrument Company
6
Keys command:
1. Press “SET” to select
2. Press return key move to next operation mode
3. Up/Down arrow keys to adjust value or select type
Main Screen: SP = 0
Press return key and release
Press and hold down “SET” for 5-sec
* RTU if using Specview
** 8 if using Specview
Factory Default Settings
TDM Module
Operation Mode Select type/value Comment
r-S Run Run/Stop
SP 0 Decimal point position
LoC OFF Lock mode (lock all keys or only up/down arrow able to use)
Out1 - (read-only) Motor output %
Operation Mode Select type/value Comment
InPt v5 Input type (v5 = 0-5V, v10 = 0-10V, nA0 = 0-20mA, nA4 =
4-20mA, nV =0-50mV)
tP-H 2206 Upper-limit range
tP-L 0 Lower-limit temperature range
CTRL PID Control mode (ON/OFF, MANUAL, PID and PID PROG)
S-HC Heat Heat/Cool control
ALA1 0 Alarm mode
SALA OFF System alarm
CoSH ON Communication write function that able to use set point from
software
C-S1 ASCII* Format type
C-no 2 Communication address
bPS 9600 Communication baud rate
Len 7** Data length setting
PrtY Even Parity bit setting
StoP 1 Stop bit setting