74 Calibration
The top and bottom portions of the Calibration dialog box
Sensor Field Highlight this field and press or to cycle through the
calibratable sensors connected to the GLX.
Measurement Field Highlight this field and press or to cycle through
the calibratable measurements of the selected sensor.
Single/All Field Highlight the Single/All Field and press to switch
between “Calibrate selected measurement only” and “Calibrate all similar mea-
surements.”
30
Calibration Type In most cases, it is not necessary to access this field because
the GLX automatically selects the most typical calibration type for the selected
measurement.
To override the automatic selection, highlight Calibration Type and press to
open the menu. Highlight the desired calibration type and press , or press the
number on the keypad corresponding to the desired type. The options are 2 Point,
1 Point Slope, and 1 Point Offset.
Pt 1 Standard Value Highlight this field and press . Enter a value and
press .
Pt 1 Sensor Input Value It is usually not necessary to access this field
because its value is entered automatically when you press .
To enter a value manually, highlight the field and press . Enter a value and
press .
Pt 2 Standard Value Highlight this field and press . Enter the value of the
standard and press .
Pt 2 Sensor Input Value It is usually not necessary to access this field
because its value is entered automatically when you press .
To enter a value manually, highlight the field and press . Enter the value of
the standard and press .
Pt 1 Standard Value
Pt 1 Sensor Input Value
Pt 2 Standard Value
Pt 2 Sensor Input Value
Arrow indicating that
additional fields can be
displayed with the up
arrow key
Sensor Field
Measurement Field
Single/All Field
Calibration Type
Arrow indicating that
additional fields can be
displayed with the down
arrow key
+
+
30
See “One-Point Offset Calibration or
Multiple-Measurement Alignment Proce-
dure” on page 77 for more information.