8. Standard-M1 (PCE-TDS 100-2)
9. Plug-in Type C
10. Standard-HS (PCE -TDS 100-3)
11. Standard-HM (PCE-TDS 100-4)
12. Standard-S1 (PCE-TDS 100-1)
13. π-Pipe
14. Standard-L1
15. Clamp-On TL-1
16. Standard-M
M24
Select sensor installation:
0. V method, 1. Z method, 2. N method, 3. W method
M25
View distance between sensors; should be stuck to as exactly as possible
M26
Save parameters to internal memory (18 memory locations)
M27
Read out saved parameters
M28
Select whether or not device should hold last good value when signal is
insufficient; standard setting is YES
M29
Low limit signal strength (default 35)
M30
Select unit system: metric or English
M31
Select flow rate unit:
Cubic Meters [m³]
Liters [l]
USA Gallons [gal]
Imperial Gallons [igl]
USA M Gallons [mgl]
Cubic Feet [cf]
USA Barrels [bal]
Imperial Barrels [ib]
Oil Barrels [ob]
Time can be per day, per hour, per minute or per second, which means 36
different unit combinations are possible
M32
Select operating unit of totaliser (options as in M31)
M33
Multiplier totaliser (default 1)
M34
Enable / disable net totaliser
M35
Enable / disable positive totaliser
M36
Enable / disable negative totaliser
M37
1. Reset totaliser
2. Reset device to factory default settings by pressing Reset key, followed by
Back key. Be careful with this function and note down your personal
settings before using it
M38
Start or stop totaliser by means of ENTER key
M39
Select the language (English or French)
M40
Set damping which can be a value between 0 and 999 seconds; if “0“ is
selected, damping is disabled
M41
Set low cutoff value
M42
Set zero point; make sure no liquid passes through the pipe
M43
Delete zero point, reset to default zero point
M44
Set flow value manually (offset value); this value should normally be “0“
M45
Set scale factor; this value is set to the delivered sensors by PCE before
shipping and should only be changed after a calibration by PCE
M46
View network IDN
M47
Lock device; parameters cannot be changed anymore
M48
Not in use
M49
Comm tester
M50
Enable or disable data logging function