5.3  Sensors 
Selection of sensor position 
The first step before installation should be finding a suitable position to place the sensors. This is a 
requirement  for  accurate  measurement  results.  Some  basic  knowledge  about  the  pipes  /  the 
plumbing system is necessary. 
The ideal location would be an infinitely long, straight pipe, whereas there must be no entrapped air 
(air bubbles) in the liquid. The pipes can either run vertically or horizontally.  To avoid inaccuracies 
due to turbulence in the liquid, a straight flow-calming section before and behind the measuring point 
should be considered. In general, the section in front of the measuring point should be at least 10 x 
the pipe diameter and after the measuring point, it should be 5 x the pipe diameter. 
The following chart shows examples of good positions: 
Pipe routings and sensor position 
Upstream 
Downstream 
L
up
 x ø 
L
dn
 x ø 
10D 
5D 
10D 
5D 
10D 
5D 
12D 
5D 
20D 
5D 
20D 
5D 
30D 
5D 