Instruction Manual

CAUTION!
Note on using the supplied safety test leads according the IEC /
EN 61010-031:2008:
Measurements in the field of overvoltage category CAT I or CAT II
can be performed with test leads without sleeves with a maximum of
up to 18mm long, touchable metallic probe, whereas for
measurements in the field of overvoltage category CAT III or CAT IV
test leads with put on sleeves, printed with CAT III and CAT IV must
be used, and therefore the touchable and conductive part of the
probes have only max. 4mm of length.
-27-