User Manual
39
3.4. DC Current
Range
Resolution
Load Voltage
Accuracy
200 µA
(< 20 µA)
0.01 µA
0.125 mV/µA
2,5% + 25 dgt.
200 µA
(> 20 µA)
0.01 µA
0.125 mV/µA
1,5% + 25 dgt.
2000 µA
0.1 µA
125 mV/µA
0,8% + 10 dgt.
20 mA
(< 2 mA)
0.001 mA
3.75 mV/µA
2,5% + 25 dgt.
20 mA
(> 2 mA)
0.001 mA
3.75 mV/µA
1,5% + 25 dgt.
200 mA
0.01 mA
3.75 mV/µA
1,2% + 10 dgt.
2 A
0.0001 A
37.5 mV/µA
2,0% + 25 dgt.
10 A
0.001 A
37.5 mV/µA
Overload protection:
0,2A / 1000V: 6,3 x 32 mm fuse in mA-Input
10A / 1000V: 10,3 x 38 mm fuse in 10A-Input
20A for max. 10 sec. every 15 min.
3.5. AC Current
Range
Resolution
Load Voltage
Accuracy
200 µA
0.01 µA
0.125 mV/µA
± 1,5% + 15 dgt.
2000 µA
0.1 µA
125 mV/µA
20 mA
0.001 mA
3.75 mV/µA
200 mA
0.01 mA
3.75 mV/µA
± 2,0% + 5 dgt.
2 A
0.0001 A
37.5 mV/µA
± 3,0% + 10 dgt.
10 A
0.001 A
37.5 mV/µA
Accuracy applies to: 10% to 100% of Range
Overload protection
0,2A / 1000V: 6,3 x 32 mm fuse in mA-Input
10A / 1000V: 10,3 x 38 mm fuse in 10A-Input
20A for max. 10 sec. every 15 min.
Frequency - range: 40 ... 1 kHz
Crest factor: CF≤3, when CF≥2, add 1% error of reading