User Manual

-30-
Resistance measurement
0Ω - 1999Ω
Resolution
1Ω
Accuracy
± (5% + 10 dgt.) at 20°C
Temperature coefficient
± 5 dgt./ 10K
Test current
30µA
Single pole phase test
100V 1000V AC
Frequency range
50Hz 400Hz
Rotary field indication
Voltage range
100 … 1000V
Frequency range
50/60Hz
Measurement principle
Double pole and contact
electrode
Safety standards
EN 61243-3:2014
Agency approvals
TÜV Rheinland - GS
Overvoltage protection
1000V AC/DC
Measurement category
CATIII 1000V / CATIV 600V
Protection degree
IP 64
Power supply
2 x 1,5V ‘‘AAA“ Batteries
Power consumption
Max. 30mA
Temperature range
-10 °C up to 55°C
Humidity
Max. 85% relative humidity