User Manual

Set the network parameters of the oscilloscope.
In the oscilloscope, press the Utility button and press H1 button, turn the M knob to select the LAN Set; press
the H2 button, the set menu is displayed on the right. Set the IP and the Port to the same value as the
“Ports-settings” in the software in step (3). Press the H3 button, it prompts “reset to update the config”. After
resetting the oscilloscope, if you can get data normally in the oscilloscope software, the connection is
successful.
Fig. 62
How to connect with the computer through a router:
1. Connection. Use a LAN line to connect the oscilloscope with a router, the LAN port of the oscilloscope is
in the right side panel; the computer should be connected to the router too.
2. Set the network parameters of the computer. Since the oscilloscope can not support obtaining an IP
address automatically, you should assign a static IP address. The Default gateway should be set
according to the router. Here we set the IP address to 192.168.1.71, Subnet mask is 255.255.255.0,
Default gateway is 192.168.1.1.
-142-