User Manual
112
How to Set the Vertical System
The VERTICAL CONTROLS includes 5 menu buttons such as CH1 ~
CH4 and Math, and 8 knobs such as VERTICAL POSITION, VOLTS/DIV
for each channel.
Setting of CH1 ~ CH4
Each channel has an independent vertical menu and each item is set
respectively based on the channel.
To turn waveforms on or off (channel, math)
Pressing the CH1 ~ CH4 or Math buttons have the following effect:
• If the waveform is off, the waveform is turned on and its menu is
displayed.
• If the waveform is on and its menu is not displayed, its menu will be
displayed.
• If the waveform is on and its menu is displayed, the waveform is turned
off and its menu goes away.
The description of the Channel Menu is shown as the following list:
Function Menu
Setting
Description
Coupling
DC
AC
Ground
Pass both AC and DC components of the input signal.
Block the DC component of the input signal.
Disconnect the input signal.
Impedance
1MΩ
50Ω
Select the input impedance to use. When 50Ω is
selected, AC in coupling is not available. If the
impedance is set to 50Ω when using AC coupling, the
coupling will be set to DC automatically.
Inverted
ON
OFF
Display inverted waveform.
Display original waveform.
Probe
X1
X10
X100
X1000
Match this to the probe attenuation factor to have an
accurate reading of vertical scale.
Limit
Full
20M
Get full bandwidth.
Limit the channel bandwidth to 20MHz to reduce
display noise.