User Manual

142
Function
Menu
Setting
Description
Type
OFF
Voltage
Time
Turn off the cursor measurement
Measure voltage using cursor
Measure time using cursor
Source
////
○,M
Select the source that the cursor measurement will be
applied to
Line
L1
Turn the G knob to adjust the position of L1
L2
Turn the G knob to adjust the position of L2
L1&L2
Two cursors are linked. Turn the G knob to adjust the
two cursors together
Perform the steps below for the voltage cursor measurement of the channel
CH1:
1. Press the Cursor button.
2. Select Source in the bottom menu; select in the left menu.
3. Select Type in the bottom menu; select Voltage in the right menu. Two
red dotted lines displayed along the horizontal direction of the screen,
which indicating L1 and L2.
4. Select Line in the bottom menu; select L1 in the right menu. The
icon will appear beside the dotted line of L1, represents you can turn
the G knob to adjust the position of L1. Select L2 to adjust the position
of L2. Select L1&L2 to adjust the two cursors together. The cursor
measure window at the top right shows the present position of two
cursors and the absolute value of voltage difference (see Figure 0-10).
Figure 0-10 Waveform of Voltage Cursor Measurement
Carry out the steps below for the time cursor measurement of the channel