This menu is only for P1360 series.
When the sample rate <=250MS/s, ADC resolution is set to 12-bit by default;
When the sample rate >250MS/s, and both channels are turned on, ADC resolution is
set to 8-bit by default;
When the sample rate <250MS/s, and only one channel is turned on, ADC resolution
can be set to 8-bit or 12-bit in this menu.
How to Set the Display System
Push the Display button and the Display menu is shown as follows:
Persist
When the Persist function is used, the persistence display effect of the picture tube
oscilloscope can be simulated. The reserved original data is displayed in fade color
and the new data is in bright color.
(1) Push the Display button.
(2) Select Persist in the bottom menu.
(3) In the left menu, turn the M knob to select the persist time, including OFF,
1second, 2second, 5second and Infinity. When the "Infinity" option is set for
Persist Time, the measuring points will be stored till the controlling value is
changed. Select OFF to turn off persistence and clear the display.
(4) Select Clear in the right menu to erase the results of previous acquisitions from
the display. The oscilloscope will start to accumulate acquisitions again.
XY Format
Function Menu
Setting
Description
Type
Dots
Vect
Only the sampling points are displayed.
The space between the adjacent sampling
points in the display is filled with the vector
form.
Persist Time
OFF
1 second
2 seconds
5 seconds
Infinity
Turn the M knob to set the persistence time
XY Mode
ON
OFF
Turn on/off XY display function
Counter
ON
OFF
Turn on/off counter
Clear
Erase the results of previous acquisitions from
the display. The oscilloscope will start to
accumulate acquisitions again.