Datasheet
4.Advanced User Guidebook
85
Source CH1 to CH4
Display the channel to which the cursor
measurement will be applied.
Perform the following operation steps for the time and voltage cursor measurement
of the channel CH1:
1. Push Cursor to display the cursor menu.
2. Select Source in the bottom menu, select CH1 in the right menu.
3. Select the first menu item in the bottom menu, the Type menu will display at
the right of the screen. In the right menu, select Time&Voltage for Type,
two blue dotted lines displayed along the horizontal direction of the screen,
two blue dotted lines displayed along the vertical direction of the screen.
Cursor measure window at the left bottom of the screen shows the cursor
readout.
4. In the bottom menu, select Line Type as Time to make the vertical cursors
active. If the Line in the bottom menu is select as a, turn the M knob to move
line a to the right or left. If b is selected, turn the M knob to move line b.
5. In the bottom menu, select Line Type as Voltage to make the horizontal
cursors active. Select Line in the bottom menu as a or b, turn the M knob to
move it.
6. Push the Horizontal HOR button to enter wave zoom mode. In the bottom
cursor menu, select Window as Main or Extension to make the cursors
shown in the main window or zoom window.
Figure 4-20 Time&Voltage Cursor Measurement
Auto Cursor
For the AutoCursr type, the horizontal cursors are set as the intersections of the
vertical cursors and the waveform.