User Manual

: Indicates the battery power is
weakening
: Beeper tone indicator for tolerance
mode
MK: Resistance (Ohm) annunciator
µmH: Inductance (Henry) annunciator
mµnpF: Capacitance (Fara) annunciator
(1) Auto Power-Off
If unused for about 10 minutes, the meter will power-down
automatically. Press Power-button switches to resume power-on
mode. When the power is down, press Power-button to turn on the
meter. The operating condition return to what they were before the
power was last turned off. In the MIN MAX record mode, RS-232
communication mode or using DC power adapter auto-power down
function is disabled automatically.
Continuous Measurement
In the power down mode, push Power-button 2 sec. Until the APO
OFF annunciator appears. Will put the meter into the continuous
measurement mode.
Power Switch
The Power-button turns the meter on or off. In the micro-computer
failure status press Power-button until the meter off.
(2) Frequency Select
Set the "FREQ"-button switch to 120 Hz or 1 kHz according to the
specimen to be test. Generally, the electrolytic capacitor is set to
120 Hz. Others are set to 1 kHz general.
(3) Parallel / Series mode
Set the "PAL SER" button switch to parallel or series measuring
circuit mode.
However, in the specimen having a high impedance in general,
measurement is made in parallel equivalent circuit mode "PAL",
while in the specimen having a low impedance, measurement is
made in series equivalent circuit mode "SER".
-39-