(11) Hi/Lo Limits
Press "Hi/Lo LIMITS" to enter the Hi/Lo LIMITS mode and change
to manual range mode. The original Hi/Lo LIMITS value and
annunciators "" "" appear in the same time individually. When
the input exceeds Hi limits, the "" is flashing and emits a
continuous tone. When the input goes below the Lo limits, the ""
is flashing and the beeper emits a pulse tone. This warns user, that
the value is out of setting ranges. Press "Hi/Lo LIMITS" again to exit
this mode. In additions, the meter will ignore and not record if
overload ("OL") situation happens in the comparative processes or
in the capacitance range 50 counts.
(12) TOL
Press "TOL" to enter the tolerance mode and change to manual
range automatically. The original preset standard value and
annunciator "TOL" appears in the same time individually. How to set
standard value, please refer to SET in this manual. When entering
TOL mode, the Main Display is the present value and the second
display the tolerance value.
There are 4 preset values in the TOL mode for instant use, just to
press "TOL" again to cycle through 1%, 5%, 10%, 20%, then back
to the present. When entering the TOL mode, the annunciators ""
and "" will appear in the same time, when the input exceeds Hi
limits, the "" is flashing and also emits a continuous tone. When
the input goes below the Lo limits, the "" is flashing and the
beeper emits a pulse tone. This warns user, that the value is out of
setting ranges.
The standard value can be set by SET function, please refer to SET
in this manual. When standard value has been set, press "TOL"
then "SET" to use preset TOL setting. In addition, the meter will
ignore and not to record if overload ("OL") situation happens in the
comparative processes or in the capacitance range 50 counts.
Press "TOL" 2 sec. to exit this function.