User Manual

1.2. Safety Symbols
The maximum current that you can measure at
this terminal is 10 A DC/AC. This terminal is
fuse protected by 10 A/1000 V fuse. When
using this range with high current, keep the
duty cycle to 30 Seconds on load and 15
Minutes off load.
The maximum current that you can measure
with this terminal is 250 mA DC/AC. This
terminal is fuse protected by 250 mA/1000
fuse.
The maximum voltage this meter can measure
is 1000 V DC or 750 V AC.
Be exceptionally careful when measuring high
voltages. Do not touch the terminals or test
lead ends!
Refer to the complete operating instructions.
Indicates protection class II, Double insulation.
CAT III Overvoltage category III
TÜV/GS approved; TÜV-Rheinland
2. Preparing of operation
2.1. Using the test leads
Use only the type of test leads supplied with your meter. These
test leads are rated for 1000 Volts.
-30-
10 A
µA/mA