User Manual
* During resistance measurement, the present measuring value
is in the main display, the measured value taken 1 second
earlier in the left secondary display, the measured value taken
2 second earlier in the middle secondary display and the
measured value taken 3 seconds earlier in the right secondary
display.
6.5. Measuring Capacitance
Warning!
When shorting capacitors (discharge), high energy discharge
may take place. Use Caution! Do not touch the terminals if there
are capacitors with voltage over 35 V DC or 25 V AC. Use
special caution in environments or rooms with adverse
environmental conditions where flammable gases, steams or
dusts may exist.
Warning!
Never attempt to measure the capacitance of a charged
capacitor, it might damage the meter.
Note:
Before measuring, give the unit about 2~3 seconds to stabilize.
1. Set the function switch to the "-II-" - position.
2. Plug the black test lead into yours meter’s COM – terminal
and the red test lead into your meter’s V/-terminal.
3. Set the RANGE-button to the correct measuring range.
4. Touch the test leads to the capacitor to be tested. The
display will indicate the proper decimal point value.
5. Read the capacitance value on the LCD-display.
6.6. Inductance Measurements
Warning!
Never attempt to measure or connect voltage, it may damage
the meter.
This function lets you measure cables or coils of 6mH to 600mH.
-41-