User Manual
Resistance ()
Range
Resolution
Accuracy
500
0,01
± 0,2% + 10 dgt.
5 k
0,0001k
± 0,2% + 5 dgt.
50 k
0,001 k
500 k
0,01 k
± 0,5% + 5 dgt.
5 M
0,0001 M
50 M
0,001 M
± 2,0% + 10 dgt.
*When using the relative mode (REL Q) to compensate for offsets.
Temperature
(type-K)
Range
Resolution
Accuracy
-200 ~ 1350
°C
0,1 °C
± 1,0% + 3,0 °C
-200 ~ 2462
°F
0,1 °F
± 1,0% + 5,4 °F
1. Does not include error of the thermocouple
probe.
2. Accuracy specification assumes ambient
temperature stable to ± 1,0 °C
Capacitance
5 nF *
0,001 nF
± 1,5% + 20 dgt.
50 nF
0,01 nF
± 1,5% + 8 dgt.
500 nF
0,1 nF
± 1,0% + 8 dgt.
5 µF
0,001 µF
± 1,5% + 8 dgt.
50 µF
0,01 µF
± 1,0% + 8 dgt.
500 µF
0,1 µF
± 1,5% + 8 dgt.
10 mF
0,01 mF
± 2,5% + 20 dgt.
*With a film capacitor or better, using relative
mode (REL Δ) to zero residual.
Duty Cycle
0,1~99,90%
0,01 %
± 1.2% +2 dgt.
Pulse width: 100 µs – 100ms
Frequency: 5 Hz – 150 kHz
-117-