User Manual

3.6. Resistance
Range
Resolution
Accuracy
220 Ω
0,01 Ω
± (0,1% + 10 dgt.)
2,2
0,1 Ω
22
1 Ω
220
10 Ω
2,2
100 Ω
22
1 kΩ
± (1,5% + 10 dgt.)
220 MΩ
100 kΩ
± (2,5% + 5 dgt.)
The above accuracies are specified from 0% to 100% of range.
For measurements in 220Ω and 2,2kΩ-range, use Relative mode to subtract the resistance of the
test leads.
The accuracy of continuity-test and diode-test are not specified.
Continuity tests:
- When the resistance is <30Ω, the buzzer will sound.
- When the resistance is > 50Ω, the buzzer will not sound.
3.7. Capacitance
Range
Resolution
Accuracy
22 nF
0,01 nF
± (2% + 5 dgt.)
220 nF
0,1 nF
2,2 µF
1 nF
22 µF
10 nF
± (2,5% + 5 dgt.)
220 µF
0,1 µF
2,2 mF
1 µF
± (5% + 5 dgt.)
22 mF
10 µF
220 mF
100 µF
unspecified
Note:
1. For film capacitors (or better), accuracy is guaranteed from 0% to 100% of range.
2. For measurements < 2pF, accuracy is guaranteed, if Relative mode (REL) is used to subtract
residual from measurements.
3.8. Logic-Frequency
Frequency Range
Sensitivity
Accuracy
20 Hz ~ 220 MHz
3 V
pp
square wave
± (0,06% + 10 dgt.)
Note:
Measurement can be up to 220MHz, but please pay attention to the signal attenuation caused by the
test leads and probes when measuring high frequency signals above 2,2MHz.
-26-