Datasheet
In the relative mode the value shown on the LCD is always The
difference between the stored reference value and the present
reading. For example, if the reference value is 24.000 V and the
present reading is 12,50 V the display will indicate –11,50 V. If the
new reading is the same as the reference value, the display will be
zero. This feature also is made as DCA ZERO adjustment.
Transformer jaws
Pick up the AC or DC current flowing through the conductor.
Trigger
Press the lever to open the transformer. When the lever is
released, the jaws will close again.
6. How to make measurements
Before making any measurements read safety precautions.
Always examine the instrument and accessories used with the
instrument for damage, contamination (excessive dirt, grease, etc)
and defects. Examine the test leads for cracked or frayed
insulation and make sure the lead plugs fit snugly into the
instrument terminals. If any abnormal conditions exist, do not
attempt to make any measurements.
6.1. Voltage measurements
1. Turn off power to the device under test and discharge all
capacitors.
2. Plug the black test lead into the COM input jack on the meter
and connect the test lead tip to a grounded point (the
reference point for measurement of voltage).
3. Select the voltage measurement position (V ) with the
function switch.
4. Select AC or DC voltage by pressing the MODE-button.
-29-