User Manual

Temperature
Range
Resolution
Accuracy
-20... 0°C 0,1°C ± 6,0% rdg. + 5°C
0 ... 400°C 0,C ± 1,5% rdg. + 4°C
400 ... 1000°C C
± 1,8% rdg. + 5°C
± 6,0% rdg. + 9°F
-4.... 32°F
0,1°F
32 ... 752°F 0,1°F ± 1,5% rdg. + 7,2°F
752 ... 1832°F F ± 1,8% rdg. + 9°F
Temperature Sensor: Type-K thermocouple - Nickel
Chromium/ Nickel Silicon
Overload protection: 600 V AC/DC
Note: Accuracy does not include error of the
thermocouple probe.
Note: Don´t use the type k thermocouple
supplied with the meter to measure
temperature above 230°C. Accuracy
does not include error of the
thermocouple probe. Accuracy
specification assumes ambient
temperature is stable to +/-1°C. For
Ambient temperature changes of +/-5°C,
rated accuracy applies after 1 hour. The
meter’s operating temperature must be
between 18°C and 28°C, otherwise
measurement accuracy is not
guaranteed. Please insure that the fuse
is OK before testing of thermocouple
-55-