User Manual
8.14. Using the Offset to Adjust for Probe Errors
Use the offset option in Setup to adjust the thermometer’s readings to
compensate for the errors of a specific the thermocouple and IR
temperature. The allowable adjustment range is ± 5.0
0
C or ± 9.0
0
F.
* Plug the thermocouple into the input connector.
* Place the thermocouple in a known, stable temperature environment
(such as an ice dry well calibrator).
* Allow the readings to stabilize.
* In Setup change the offset until the primary reading matches the
calibration temperature.(“Changing Setup Options.")
8.15. Battery Replacement
Power is supplied by a 9 V battery (NEDA 1604, IEC 6F22). The
battery sign appears on the LCD when replacement is needed. To
replace the battery, remove the battery compartment from the back of
the meter and lift off the battery cover. Remove the battery from its
contacts.