User Manual

4. As soon as the first measurement is completed, press the
AVG button (you will hear a single beep) and then release, the
display will show a “HOLD” and a “AVG” under the reading.
The reading is now being moved to internal memory and the
unit will display the average reading and number of points
measured. The average reading is held for about 3 seconds.
Repeat steps 4 until all desired points have been measured.
This unit can record up to 20 points at one time. In AVG
mode, press and hold AVG button for about 3 seconds to
return and press AREA button to clear memory of multi-point
average reading.
5. The Meter is now ready to measure air flow average for
multipoints.
5.5. Area set
1. Power unit on.
2. Press UNITS key and select CFM or CMM.
3. Hold AREA key until beeps twice, the “AREA” will display
and enter area set mode.
4. Press the MAX/MIN key to shift the radix point.
5. Press the HOLD key to shift the flash position.
6. Press the UNITS key to adjust digit which is flash.
NOTES:
The CFM/CMM-measurement is for measuring the air flow volume
of a certain area.
For example: Measuring air flow volume of an air-condition
Set the area with value of the air output area of the aircondition.
During measurement the PeakTech
®
5060 will calculate and show
automatically the air-flow volume in reference to the set area value
in the display.