Service Manual

12
REAL™ Immersive System is intended for use in the electromagnetic environment specified below. The customer or the
user of REAL Immersive System should assure that it is used in such an environment.
Emissions Test
Compliance
RF emissions
CISPR 11
Group 1
REAL Immersive System uses RF energy only for
its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interfer-
ence in nearby electronic equipment.
RF emissions
CISPR 11
Class A
REAL Immersive System is suitable for use in all
establishments other than domestic and those
directly connected to the public low-voltage power
supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage Fluctuations/
Flicker emissions
Complies
REAL Immersive System is intended for use in the electromagnetic environment specified below. The customer or the user of
REAL Immersive System should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic
environment guidance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
(6 Vrms in ISM radio Bands
within 150kHz 80MHz)
3 Vrms
Portable and mobile RF communications
equipment should be used no closer to any
part of REAL Immersive System, including
cables, than the recommended separation
distance calculated from the equation appli-
cable to the frequency of the transmitter.
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.7 GHz
3 V/m
Recommended separation distance
d = 1.2√P
d = 1.2√P 80 MHz to 800 MHz
d = 2.3√P 800 MHz to 2.7 GHz
where P is the maximum output power rating
of the transmitter in watts (W) according
to the transmitter manufacturer and d is
the recommended separation distance in
meters (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey
a
, should be less than the compliance
level in each frequency range
b
.
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflec-
tion from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radi-
os, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess
the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which REAL Immersive System is used exceeds the applicable RF compliance
level above, REAL Immersive System should be observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as reorienting or relocating REAL Immersive System.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
REAL Immersive System is intended for use in the electromagnetic environment specified below. The customer or the user
of REAL Immersive System should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic
environment guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
Floors should be wood, concrete
or ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
2 kV for
power supply lines
±1 kV for
input/output lines
2 kV for
power supply lines
±1 kV for
input/output lines
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
±1 kV differential mode
±2 kV common mode
1 kV differential mode
±2 kV common mode
Mains power quality should be that
of a typical commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Voltage Dips 30% reduction,
25/30 periods
At 0°
Voltage Dips 30% reduction,
25/30 periods
At 0°
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the
EQUIPMENT requires continued
operation during power mains
interruptions, it is recommended that
REAL Immersive System be powered
from an uninterruptible power supply
or a battery.
Voltage Dips > 95% reduc-
tion, 0.5 period
At 0°, 45°, 90°, 135°, 180°,
225°, 270° and 315°
Voltage Dips > 95% reduc-
tion, 0.5 period
At 0°, 45°, 90°, 135°, 180°,
225°, 270° and 315°
Voltage Dips > 95% reduc-
tion, 1 period At 0°
Voltage Dips > 95% reduc-
tion, 1 period At 0°
Voltage Interruptions > 95%
reduction, 250/300 periods
Voltage Interruptions > 95%
reduction, 250/300 periods
(50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commer-
cial or hospital environment.
Recommended separation distances between
portable and mobile RF communications equipment and REAL Immersive System
REAL Immersive System is intended for use in an electromagnetic environment in which radiated RF disturbances are con-
trolled. The customer or the user of REAL Immersive System can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications equipment (transmitters) and REAL Immersive System
as recommended below, according to the maximum output power of the communications equipment.
Rated maximum
output power
of transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
d = 1.2√P
80 MHz to 800 MHz
d = 1.2√P
800 MHz to 2.7 GHz
d = 2.3√P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can
be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of
the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflec-
tion from structures, objects and people.