Commissioning
2022-11
35
Single Read Special Read-Only Code
As confirmation, read out the read-only code of the tag within the read/write head's detection
range via the single read special read-only code command. See "Single Read Special Read-
Only Code SS" on page 38.
Parameterizing the Read/Write Head
Requesting and Setting the Transmission Power
Read out the read/write head's transmission power with the read parameter PT command:
The read/write head's set transmission power is 50 mW (32
hex
corresponding to 50
dec
).
Change the transmission power of the read/write head to 100 mW (100
dec
corresponding to
64
hex
) via the write parameter PT command:
Serial Ethernet PROFIBUS/PROFINET
Command: SS10 .00.04.0A.02 .0A.02
Confirmation: - .00.06.0A.02.FF.30 .0A.02.FF.30
Response: .30.31.00.0E.30.
00.30.14.F7.33.7
C.00.1F.00.00.03
.1C.6E
.46.31.30.30.30.3
1
.00.16.0A.02.00.31.00.0E.
30.00.30.14.F7.33.7C.00.1
F.00.00.03.1C.6E
.00.0A.0A.02.0F.32.30.30.
30.31
.0A.02.00.31.00.0E.30.00.
30.14.F7.33.7C.00.1F.00.0
0.03.1C.6E
.0A.02.0F.32.30.30.30.31
Serial Ethernet PROFIBUS/PROFINET
Command: RP1UPT.00.00 .00.0A.BE.03.00.55.50.54.
00.00
.BE.03.00.55.50.54.00.00
Confirmation: - .00.06.BE.03.FF.33 .BE.03.FF.33
Response: .30.31.00.32 .00.0A.BE.03.00.34.00.02.
00.32
.BE.03.00.34.00.02.00.32
Serial Ethernet PROFIBUS/PROFINET
Command: WP1UPT.00.02.0
0.64
.00.0C.BF.03.00.55.50.54.
00.02.00.64
.BF.03.00.55.50.54.04.00.
02.00.64
Confirmation: - .00.06.BF.03.FF.35 .BF.03.FF.35
Response: .30.31 .00.06.BF.03.00.36 .BF.03.00.36