3.0 Installation
Mounting
The Micro-GASS enclosure is rated NEMA-4 however, the system is intended for indoor
installation. When choosing a mounting location, choose a location that avoids exposure to
direct sunlight as it may raise the internal temperature beyond the system operating limits. The
system should also not be subjected to extremely cold temperatures. Wall mounting of the
system is required for proper operation of the flow meter. A set of four mounting feet are
supplied for mounting the system on 12.4” centers.
Figure 2 – Mounting Dimensions
Electrical Connections
Confirm the operating voltage on the system nameplate before connecting power.
Electrical conduit connection -
The system is supplied via the ½” electrical hub that can
be used to attach sealtite™ flexible conduit or rigid metal or plastic conduit
.
Electrical conductor connections -
Made at a terminal block located inside the
enclosure near the conduit hub. The MicroGASS system is factory configured for either 115 or
220VAC operation.
Heated Line connection – A three pole Molex style inline connector is provided to supply
10A fused power to a heated line if desired. Voltage will be present at this connector whenever
power is connected to the system. For this reason, it is recommended that the system power be
supplied via a fused disconnect switch.