PERMA PURE LLC Mini-GASS™ Sample Conditioning System User’s Manual PERMA PURE LLC 8 Executive Drive ▪ P.O. Box 2105 ▪ Toms River, NJ 08754 (732) 244-0010 ▪ (800) 337-3762 ▪ fax (732) 244-8140 ▪ www.permapure.com ▪ info@permapure.
PERMA PURE LLC Mini-GASS™ User’s Manual Doc. #212: Revision: 00A Page 2 of 32 Contents Unpacking.............................................................................................................3 Important safety warnings ......................................................................................5 General description ................................................................................................6 Specifications and Features ...........................................
PERMA PURE LLC Mini-GASS™ User’s Manual Doc. #212: Revision: 00A Page 3 of 32 Unpacking Perma Pure has made every effort to ship you a high quality product that has been thoroughly inspected and tested. It has been carefully packed to ensure that it arrives at your facility in good condition. Even though every effort has been made to prevent damage during transportation, damage can occur by the carrier.
PERMA PURE LLC Mini-GASS™ User’s Manual Doc.
PERMA PURE LLC Mini-GASS™ User’s Manual Doc. #212: Revision: 00A Page 5 of 32 Important safety warnings Please be sure to review the following basic safety procedures. These procedures represent the MINIMUM requirements to operate the equipment safely. It is the ultimate responsibility of the operator to ensure proper safety practices are utilized at the point of operation. • This equipment is NOT designed to operate in a wet environment. • Condensate is potentially dangerous.
PERMA PURE LLC Mini-GASS™ User’s Manual Doc. #212: Revision: 00A Page 6 of 32 General description Proven tube-in-shell membrane technology is the basis of the Mini-GASS™ Conditioning System, which selectively removes water vapor from a gas stream. The driving force is the vapor pressure differential between the sample gas and a purge gas counter-flowing outside the tubing. This difference quickly dries the gas stream.
PERMA PURE LLC Mini-GASS™ User’s Manual Doc. #212: Revision: 00A Page 7 of 32 Specifications and Features Sample flow rate 0 to 10 lpm Sample inlet temperature 250°F (121°C) max. with stainless steel filter 230°F (110°C) max.
PERMA PURE LLC Mini-GASS™ User’s Manual Doc. #212: Revision: 00A Page 8 of 32 Installation Mounting 1. Install Mini-GASS system on vertical surface with dryer/filter compartment on top and control compartment on bottom. 2. Place mounting feet on each corner of enclosure with slotted end protruding at a 45° angle. 3. Drop mounting screw in from top and tighten into foot. Connections (Refer to figure 2) Heated sample line HUB FITTING 1. Install heated line sealing fitting by threading hub into sleeve. 2.
PERMA PURE LLC Mini-GASS™ User’s Manual Doc. #212: Revision: 00A Page 9 of 32 Heatless air dryer (optional) 1. Connect oil-free compressed air line to port labeled “Purge Gas Inlet” 2. A ¼” FNPT bulkhead fitting on bottom of enclosure is the heatless dryer purge air exhaust, see figure 3. Humid air can be vented to atmosphere or piped to a remote location. Purge air exhaust (optional) To vent the purge gas exhaust to a remote location, connect a line to the ¼” FNPT fitting labeled “Purge Exhaust Port”.
PERMA PURE LLC Mini-GASS™ User’s Manual Doc. #212: Revision: 00A Page 10 of 32 used for runs up to 10 feet, or larger i.d. tubing for longer exhaust lines. Line should not restrict purge flow. Purge eductor inlet connection (optional) Connect air supply line to compression fitting labeled “Instrument Air”. Steam connection (optional) (Refer to figure 4) 1. Insulate supply line to steam control valve. 2. Insulate supply line between valve and steam coil. 3.
PERMA PURE LLC Mini-GASS™ User’s Manual Doc. #212: Revision: 00A Page 11 of 32 the equipment, within easy reach of the operator and shall be clearly marked as the disconnecting device for the Mini-GASS. Connect power supply line to terminal block located on the backplate of the control enclosure inside the ½” conduit hub. (Refer to figure 5 for color coding). Connect the system to an appropriate earth ground. Adhere to all electrical code requirements in effect at the installation site.
PERMA PURE LLC Mini-GASS™ User’s Manual Doc. #212: Revision: 00A Page 12 of 32 Automatic Filter Drain Control (Optional) Automatic filter drain is controlled by a repeat cycle timer that operates a solenoid valve. Two DIP switches are located on timing device to control drain and cycle times located on backside of control panel. Standard factory setting is to drain for 0.1 minute once every 24 hours. Drain Time (left DIP switch) Cycle Time (right DIP switch) 0.
PERMA PURE LLC Mini-GASS™ User’s Manual Doc. #212: Revision: 00A Page 13 of 32 Steam Temperature Control Steam heated systems are regulated by a thermostatic steam control valve located in the purge air exhaust stream. This valve is preset at the factory to an appropriate temperature determined by sample conditions. A typical setpoint temperature is 180° F which is the approximate temperature of the purge air exhaust indicated by analog thermometer located in the purge air exhaust stream.
PERMA PURE LLC Mini-GASS™ User’s Manual Doc. #212: Revision: 00A Page 14 of 32 System fine tuning Temperature adjustment Steam heated systems Steam control valve is factory set to 180° F and is normally not field adjustable. This temperature is adequate for most applications in which sample gas water content is not in excess of 50% water vapor by volume. Consult factory if condensation in sample tubing is detected. Electrically heated systems 1.
PERMA PURE LLC Mini-GASS™ User’s Manual Doc. #212: Revision: 00A Page 15 of 32 6. Place new element in groove in bottom piece. Be sure that element is seated correctly and parallel to glass shell. 7. Carefully mate bottom assembly onto top piece. Slight twisting motion may be required to allow shell to slip over o-ring seal. 8. Visually make sure element is seated correctly in top groove. 9. Replace bolt through hole in bottom piece and screw clockwise into top piece. Do not over-tighten center bolt.
PERMA PURE LLC Mini-GASS™ User’s Manual Doc. #212: Revision: 00A Page 16 of 32 3. Separate housing and bottom cap as an assembly from top cap. 4. Remove spring and top screen. 5. Remove old media and dispose of properly (rinse housing with soapy water to clean). 6. Fill housing with 50cc of Berl Saddles (tap housing to allow material to settle). 7. Pour 150cc of Scrubbing Media (tap housing to allow material to settle). 8. Replace stainless steel screen on top of media. 9. Replace spring on top of screen.
PERMA PURE LLC Mini-GASS™ User’s Manual Doc.
PERMA PURE LLC MG-PUMPKIT fillings Pump repair kit, Teflon diaphragm SP-270E Ceramic filter element for MG-1228P Mini-GASS™ User’s Manual Doc.
PERMA PURE LLC Mini-GASS™ User’s Manual Doc. #212: Revision: 00A Page 19 of 32 APPENDIX A: Warranty and disclaimers Perma Pure LLC Perma Pure (Seller) warrants that product supplied hereunder shall, at the time of delivery to Buyer, conform to the published specifications of Seller and be free from defects in material and workmanship under normal use and service.
PERMA PURE LLC Mini-GASS™ User’s Manual Doc. #212: Revision: 00A Page 20 of 32 APPENDIX B: Dryer element replacement TO DISASSEMBLE DRYER Tools Needed: - Phillips head screwdriver - unsharpened pencil with eraser 1. Loosen locking screws on both ends of dryer. 2. Insert eraser end of pencil into one sample port until it rests on tube header face (Refer to Figure 8). 3. Hold dryer vertically and place other end of pencil down onto a hard, slip resistant surface. 4.
PERMA PURE LLC Mini-GASS™ User’s Manual Doc. #212: Revision: 00A Page 21 of 32 Figure 10 APPENDIX C: Flange Mounted with Integral Sampling Probe and Filter Blowback Installation Flange mounted Mini-GASS® is designed to be stack mounted on 2", 3" or 4" pipe flange. The integral sample probe and filter eliminates need for a heated umbilical line to connect stack gas sampling probe to sample conditioning system. The Mini-GASS system is attached directly to sample port flange by following the steps below: 1.
PERMA PURE LLC Mini-GASS™ User’s Manual Doc. #212: Revision: 00A Page 22 of 32 10. Connect free end of turnbuckle to anchor point at middle back of system (refer to Figure 14). 11. Attach with stainless clevis pin. 12. Snap keeper wire into place. 13. Adjust turnbuckle by turning it out to increase length until union fitting is straight. Turnbuckle is used only to relieve torsional load from union fitting. 14. Tighten union fitting with a wrench. 15.
PERMA PURE LLC Mini-GASS™ User’s Manual Doc. #212: Revision: 00A Page 23 of 32 2. Connect cal gas line to fitting labeled “Cal Gas Inlet”. 3. Connect 110VAC power to blow back timer Note: This also provides power to the filter in the probe. Blow Back Operation The blow back option includes a timer circuit which will initiate the blow back cycle based on a regular, adjustable time period. The time cycle control turned to the far left will cycle the blow back every 15 minutes.
PERMA PURE LLC Mini-GASS™ User’s Manual Doc.
PERMA PURE LLC Mini-GASS™ User’s Manual Doc.
PERMA PURE LLC Mini-GASS™ User’s Manual Doc.
PERMA PURE LLC Mini-GASS™ User’s Manual Doc.
PERMA PURE LLC Figure 18 Mini-GASS™ User’s Manual Doc.
PERMA PURE LLC Mini-GASS™ User’s Manual Doc.
PERMA PURE LLC Mini-GASS™ User’s Manual Doc. #212: Revision: 00A Page 30 of 32 APPENDIX E: Z-Purge Operation With the inert gas supply connected, enclosure power de-energized and alarm system (if utilized). Follow steps to purge system: 1. Carefully read Start-Up Instruction Nameplate on system. 2. Check operation of the Enclosure Protection Vent (EPV-1, if utilized), opening it manually several times (see helpful hints below) 3. Seal protected enclosures. 4.
PERMA PURE LLC Mini-GASS™ User’s Manual Doc. #212: Revision: 00A Page 31 of 32 APPENDIX F: Safety Interlock System Control system safety interlock The interlock system is present to provide an output signal that can be used internally to stop sample flow via the optional internal sample pump. This system can also be used to provide an external output to a PLC or other control/monitoring system so that the status of the Mini-GASS can be monitored.
PERMA PURE LLC Mini-GASS™ User’s Manual Doc.