PERMA PURE LLC SDS™-Series 4C-D201/-D202 Supplemental Drying System User’s Manual PERMA PURE LLC 8 Executive Drive ▪ P.O. Box 2105 ▪ Toms River, NJ 08754 (732) 244-0010 ▪ (800) 337-3762 ▪ fax (732) 244-8140 ▪ www.permapure.com ▪ info@permapure.
SDS™-Series User’s Manual Doc. #219: Revision: 000 Page 2 of 12 PERMA PURE LLC Contents Unpacking ........................................................................................................................ 3 Introduction...................................................................................................................... 3 General description ............................................................................................................
PERMA PURE LLC SDS™-Series User’s Manual Doc. #219: Revision: 000 Page 3 of 12 Unpacking Perma Pure has made every effort to ship you a high quality product that has been thoroughly inspected and tested. It has been carefully packed to ensure that it arrives at your facility in good condition. Even though every effort has been made to prevent damage during the transportation process, damage can occur by the carrier.
SDS™-Series User’s Manual Doc. #219: Revision: 000 Page 4 of 12 PERMA PURE LLC General description Proven tube-in-shell membrane technology is the basis of the SDS™-Series Drying System, which selectively removes water vapor from a gas stream. The driving force is the vapor pressure differential between the sample gas and a purge gas counter-flowed outside the tubing. This difference drives the reaction, quickly drying the gas stream.
PERMA PURE LLC SDS™-Series User’s Manual Doc. #219: Revision: 000 Page 5 of 12 Important safety warnings Please be sure to review the following basic safety procedures. These procedures represent the MINIMUM requirements to operate the equipment safely. It is the ultimate responsibility of the operator to ensure proper safety practices are utilized at the point of operation. • This equipment is NOT designed to operate in a wet environment. • Condensate is potentially dangerous.
SDS™-Series User’s Manual Doc. #219: Revision: 000 Page 6 of 12 PERMA PURE LLC Installation Tools and hardware recommended for installation: 5/8” wrench • (2) • (1) ¾” wrench • Teflon tape or pipe thread sealant • (4) ¼” dia.
PERMA PURE LLC SDS™-Series User’s Manual Doc. #219: Revision: 000 Page 7 of 12 Mounting 1. Attach system to secure wall using (4) ¼” dia. screws. 2. Refer to figure 3 for mounting dimensions.
SDS™-Series User’s Manual Doc. #219: Revision: 000 Page 8 of 12 PERMA PURE LLC Sample connections – Model SDS-4C-D201 (Refer to figure 4) 1. Connect the sample supply line to the ¼” tube compression fitting labeled “Sample gas inlet”. 2. Connect the sample outlet line to the ¼” tube compression fitting labeled “Sample gas outlet”.
PERMA PURE LLC SDS™-Series User’s Manual Doc. #219: Revision: 000 Page 9 of 12 Dual stream application (Refer to figure 5) 1. Remove the crossover tube. 2. Connect the sample stream 1 supply line to the ¼” tube compression fitting labeled “Sample gas inlet”. 3. Connect the sample stream 1 outlet line to the 1/4” tube compression fitting B. 4. Connect the sample stream 2 supply line to fitting A. 5. Connect the sample stream 2 outlet line to fitting labeled “Sample gas outlet”.
SDS™-Series User’s Manual Doc. #219: Revision: 000 Page 10 of 12 PERMA PURE LLC 2. Optional exhaust connection: To vent the purge gas exhaust to a remote location, connect a line to the ¼” FNPT fitting labeled “Purge Air Vent”. This is suggested for gas samples that are toxic, flammable or otherwise hazardous. In case of failure of seal integrity, the sample will be vented to the remote location. For runs under 10 ft, ¼” OD tube is acceptable. For runs 10-40 ft, 3/8” OD tube is recommended.
SDS™-Series User’s Manual Doc. #219: Revision: 000 Page 11 of 12 PERMA PURE LLC Troubleshooting guide The SDS-Series is a very simple component of the sample conditioning system with no electricity, heat or moving parts to fail. Problems can arise if there is a failure upstream which causes liquid carryover or a high pressure or vacuum spike in the line. If these issues are suspected, please call the factory.
SDS™-Series User’s Manual Doc. #219: Revision: 000 Page 12 of 12 PERMA PURE LLC Appendix A: Warranty and disclaimers Perma Pure LLC Perma Pure (Seller) warrants that product supplied hereunder shall, at the time of delivery to Buyer, conform to the published specifications of Seller and be free from defects in material and workmanship under normal use and service.