User's Manual

Degree of protection
against harmful ingress
of water of particular
matter
IP22
Applied part
(measurement
platform)
Type BF
Protection against
electric shock
Internally powered
The device is not suitable for use in the presence of flammable anesthetic
mixtures with air or with oxygen or nitrous oxide.
Electromagnetic emissions and immunity
The device is approved according to EMC safety standard EN 60601-1-2. It
is designed to be used in typical domestic environments.
EMC Guidance
-
The scale needs special precautions regarding EMC and needs to be
installed and put into service according to the EMC information provided
in the accompanying documents.
-
Wireless communications equipment such as wireless home network
devices, mobile phones, cordless telephones and their base stations,
walkie-talkies can affect this equipment and should be kept at least a
distance d = 3.3 m (11ft) away from the equipment.
Note: As indicated in IEC 60601-1-2:2007 for ME equipment, a typical cell
phone with a maximum output power of 2 W yields d = 3.3 m (11ft) at an
immunity level of 3V/m.
Table 1 Guidance and manufacturer's declaration –
electromagnetic emissions - for all ME equipment and ME
systems
Guidance and manufacturer’s declaration – electromagnetic emissions
The device is intended for use in the electromagnetic environment specified
below. The customer or the user of the device should assure that it is used in
such an environment.
Emissions test
Compliance Electromagnetic environment -
guidance
RF emissions CISPR 11 Group 1 The device must emit electromagnetic energy
in order to perform its intended function.
Nearby electronic equipment may be
affected.
21
English