ST/AR ST Segment Algorithm
ST/AR Arrhythmia & ST Segment Monitoring
7-11
Establishing
ST Reference
Beats
(Baseline)
After adjusting the measurement points, you can establish baseline reference
beats for all available leads in the ST Review window at the IntelliVue
Information Center. Reference beats enable you to compare waveform changes,
for example from admission, or prior to or after treatment. The reference
continues to be saved beyond the 24 hour review window, but you can update it
to any beat within the last 24 hours. Please refer to the IntelliVue Information
Center Instructions for Use or Online Help for directions.
ST Alarm
Settings
All IntelliVue Information Center alarm settings (limits and on/off status) have
unit default settings. The IntelliVue Information Center however, lets you set the
high and low ST alarm limits for individual patients based on:
Your assessment of the patient's clinical condition.
Unit protocols.
3 Adjust the J point, if necessary, by placing the cursor over the J-
point button to access the adjustment arrows. Then use the arrows to
position the bar at the end of the QRS complex and the beginning of
the ST segment.
4 Adjust the ST point, if necessary, by using the J point as an
“anchor” and placing the bar at the midpoint of the ST segment.
Choices are J+0, J+20, J+40, J+60, or J+80.
Step
Action
T
R
P
S
Q
J point
T
R
P
S
Q
ST point