Connecting the SpO
2
Cable
8-14 SpO
2
Monitoring
Connecting the SpO
2
Cable
Caution
Extension cables: The use of extension cables presents a tripping hazard. Use
appropriate care when they are in use.
Electrical Interference: Position the sensor cable and connector away from
power cables, to avoid electrical interference.
Step Action
1 Connect the sensor cable to IntelliVue TRx
+
.
• Connect reusable sensors directly into the transceiver.
• Connect disposable sensors into the adapter cable, then
connect the adapter cable to the transceiver. Remove the
protective backing.
2 Adjust SpO
2
alarms in the Patient Window (see “Telemetry
Controls in the Patient Window” on page 9-2).
3 Make other adjustments in the Telemetry Setup Window (see
“Patient Configurable Settings in Telemetry Setup” on page 9-4).