Alarm Behavior with Telemetry Overview
10-4 Pairing with IntelliVue Patient Monitors
Alarm
Suspend/
Pause
When alarms are suspended/paused, the messages and types of alarms affected
depend on where the suspend/pause was initiated.
Alarm Silence When an active alarm is silenced, the types of alarms that are silenced depend on
the alarm source and where the silence was initiated.
If alarms are
suspended/
paused
from....
these alarms are
suspended/
paused
and this message appears
Information
Center
both bedside and
telemetry
measurements
Information Center: ALARMS
SUSPENDED or ALARMS
PAUSED
Bedside: ALARMS OFF in
Overview window and ALARMS
OFF or ALARMS PAUSED on the
monitor (depending on
configuration)
IntelliVue
Patient Monitor
bedside
measurements only
Information Center: BED
ALARMS SUSPENDED or BED
ALARMS PAUSED (depending on
configuration)
Bedside: ALARMS OFF or
ALARMS PAUSED