User's Manual

Table Of Contents
Alarm Behavior with Telemetry Overview
Pairing with IntelliVue Patient Monitors
10-5
WarningWarning
If the remote Silence key in the Overview window is enabled for IntelliVue
monitors connected to the Information Center, remote silencing for these
beds may be enabled in other clinical units.
Alarm
Source
Where
Silenced
Effect at
Paired
Bedside
Effect at
Information
Center
Bedside alarm Bedside
Silence button
Alarm is
silenced
Bedside alarm is
silenced. There is
no effect on
telemetry alarms
(if Global
Silence is not
enabled at the
Information
Center)
Telemetry
alarm
Bedside No effect on
telemetry
alarms
No effect on
telemetry alarms
Bedside and/or
telemetry alarm
Telemetry Data
Window at
bedside (if
configured)
Bedside and/or
telemetry alarm
is silenced
Bedside and/or
telemetry alarm
is silenced (if
Silence
Overview
Alarms is
enabled at the
Information
Center)