Measurement Specifications
Safety Standards & Specifications
12-15
Measurement Specifications
ECG
Parameter Specification
ECG channel
transmitted Leads
• 3 electrodes
• 5 electrodes
• 5 electrodes,
EASI
• 6 electrodes
• Channel #1 = I, II, or III
• Channel #1 = II
Channel #2 = III
Channel #3 = MCL
• Channel #1 = Va-i
Channel #2 = Va-s
Channel #3 = Ve-s
• Channel #1= II
Channel #2 = III
Channel #3 = Va
Channel #4 = Vb
Resolution 5 μV
ECG Input Differential, defibrillator protected against 360
joules discharge into a 100 ohm load
Input Impedance > 5 megohms (@ 10 Hz
Input Dynamic Range +/- 9 mV
DC Offset Range +/- 320 mV
CMRR >
90 dB @ 50, 60 Hz
Bandwidth +/- 3 dB 0.05 to 40 Hz