Measurement Specifications
12-16 Safety Standards & Specifications
Gain Accuracy +/- 5% at 25
o
C (77
o
F)
Noise Referred to ECG
Input
AAMI: 30 μV
Lead Wires 3, 5 or 6-wire leadset. 5-lead compatible with
IntelliVue Patient Monitor, AAMI/IEC color
codes
Time to baseline from
Defibrillator
AAMI: 5 s max (until ECG wave is on display
but not yet centered, monitoring bandwidth)
Pacer Rejection
Performance
(Pace pulses with no
tails).
Positive pacers
1
Amplitude Width
+2 to +700 mV 0.1, 0.2, 0.5 and 1.0 ms
+2 to +500 mV 1.5 ms
+2 to +400 mV 2 ms
Negative pacers
1
Amplitude Width
-2 to -700 mV 0.1, 0.2, 0.5 and 1.0 ms
-2 to -500 mV 1.5 ms
-2 to -400 mV 2 ms
1
Philips does not claim, verify, or validate
support for all available pacemakers.
EMC Performance
Limits, radiated
immunity
Meets Essential Performance, but may observe
some waveform disturbance over 181 to 202 MHz
@ >1.8 V/m and 203 to 213 MHz @ > 1.0 V/m
Parameter Specification