User's Manual

Table Of Contents
Measurement Specifications
Safety Standards & Specifications
12-17
SpO
2
Parameter Specification
SpO
2
Measurement
Range (Calibration and
Display)
0 to 100%
SpO
2
Accuracy See table following.
SpO
2
Resolution 1%
SpO
2
Numerics -
Averaging
10 seconds
Note—The update rate for the SpO
2
pulse
oximetry value and pulse rate is typically 1
second. This can be extended to a max. 60 s
when NIBP is measured on the same limb, with a
corresponding INOP message after a max. of 30
s, indicating that the displayed values are not
current values.
The effect of SpO
2
pulse oximetry on data
averaging is internally controllable by the
transmitter, with no user controls.
SpO
2
& Pulse Numerics
- Update Rate
Transmitted once per second.
Pleth Wave- Sampling
Rate
125 sps
Technical Alarms
(INOPs)
Triggered if the sensor is disconnected, if a pulse
is not detected, if the signal is noisy, if light
interference is detected, if the sensor is defective,
if the measurement is erratic, or if the module is
malfunctioning