User's Manual

Table Of Contents
Technical Alarms (INOPs)
Alarms
5-11
NO SIGNAL Hard Patient is out of
range, or
No batteries in
transceiver, or
Transceiver has
failed, or
Connection to
Database Server has
failed.
Make sure that the
transceiver is in range
and has good batteries.
Replace the transceiver
if Power On Self Test
fails.
Put bed in Standby.
Contact Service.
REPLACE BATTERY
REPLACE BATTERY T
Note—This INOP may also
be configured to display as
a Red or Yellow Technical
Alarm.
Hard,
Latched
Dead battery. No
monitoring is occurring.
Replace batteries.
SOME ECG ALRMS OFF Soft Some yellow arrhythmia
alarms have been turned off
for this patient.
For information only.
SpO
2
T EQUIP MALF Hard Malfunction in the SpO
2
equipment
Call Service.
SpO
2
T ERRATIC Hard Erratic SpO
2
measurements, often due to
a faulty sensor or invalid
SpO
2
measurements, or
incorrect transducer
position
Repeat measurement,
reposition sensor on patient,
or finally, replace sensor.
Technical Alarms (Inoperative Conditions)
Alarm Text Priority Condition What to do