Selecting a SpO
2
Sensor
SpO
2
Monitoring 8-9
To use the chart:
Find the patient’s weight on the vertical axes. On the horizontal axis at this
weight, the shaded areas indicate that the sensor is a “best choice” for the
patient. Unshaded areas indicate a “good choice.” For example, the best
reusable sensor for a 50 kg patient is the M1191A, applied to the finger or toe.
Alternatively, you could use M1194A applied to the ear.
Selecting an SpO
2
Sensor
Preferred sensor types
Alternative sensor types
Disposable
sensor
M1192A M1191A M1194A
Philips
M1901B
N-25
N
M1903B
D-20
P
M1904B
D-25
A
110
88
66
44
33
Pedi
Adult
Patient
Weight
( kg )
50
40
30
20
15
Pedi
Adult
Adult
Pedi
Adult
Patient
Weight
( lb )
Pedi
Pedi
Pedi
Adult
Oxisensor
OxiMax
M1196A
Reusable
sensor