Transceiver Operation when Connected to TeleMon
9-16 Telemetry Functions at the Information Center & TeleMon
Transceiver Operation when Connected to TeleMon
When connected to TeleMon, the following differences in transceiver operation
will be in effect.
Important—The transceiver must have good batteries installed before being
connected to TeleMon.
Alarms
• Alarm Suspend/Pause and Standby can be initiated only from TeleMon,
not from the transceiver or the Information Center. The alarm suspend
indicator on the transceiver will accurately reflect the current state of
alarm pause.
ECG Operation
• Vb, the second V-lead in 6-wire lead-set, is not supported.
• After a change in leadset, the TeleMon returns to the default ECG
settings, and arrhythmia relearn occurs automatically. Be sure to check the
monitoring leads after you switch leadsets.
SpO
2
Operation
•SpO
2
is always in continuous mode.
• Changes in SpO
2
mode do not take effect until after the transceiver is
disconnected from TeleMon. Mode settings are defined in the following
table:
• If the sensor is disconnected from the transceiver during SpO2 monitoring
at TeleMon, a No Sensor technical alarm will be displayed at the
Information Center. NO VISUAL OR AUDITORY INDICATORS FOR
THE NO SENSOR ALARM WILL BE ACTIVE AT TELEMON.
Mode Selected at TeleMon Mode at Disconnected Transceiver
Continuous Continuous
5-min. Continuous
1-min. Continuous
Manual Spot Check