Physiologic Alarms
5-6 Alarms
** MULTI ST Lx, Ly Yellow Two ST leads (Lx and Ly) exceed alarm
limit elevation or depression for > 60
seconds (EASI mode or when selected).
ST/AR ST
** NURSE CALL Yellow The Telemetry button on the transceiver
has been pressed (and the system is
configured to alarm and the Telemetry
button is on.
Also initiated if the installation includes a
paging system and the Information Center
is configured for paging upon receipt of
Nurse Call signal.
Note—For IntelliVue Information Centers
Release J and later, Nurse Call is a
yellow, one star (*) alarm.
Clinician-
initiated at
transceiver
or by
paging
subsystem
** SpO
2
T yyy > xxx Yellow High SpO
2
.
SpO
2
value (yyy) greater than high SpO
2
limit (xxx).
SpO
2
** SpO
2
T yyy < xxxx Yellow Low SpO
2
.
SpO
2
value (yyy) less than low SpO
2
limit
(xxx).
SpO
2
** ST lead > xxx Yellow STlead is greater than the Elevation limit. ST/AR
** ST lead < xxx Yellow STlead is less than the Depression limit. ST/AR
* HR yyy > xxx Yellow Heart Rate (yyy) greater than the upper
Heart rate limit (xxx).
ST/AR
Basic &
Enhanced
Arrhythmia
* HR yyy < xxx Yellow Heart Rate (yyy) lower than the lower
Heart Rate limit (xxx).
ST/AR
Basic &
Enhanced
Arrhythmia
Physiologic (Patient) Alarms
Alarm Text Priority Condition Source