User's Manual

Table Of Contents
Technical Alarms (INOPs)
5-10 Alarms
CANNOT ANALYZE ST Soft ST algorithm cannot
reliably generate any valid
ST values on any
monitored lead.
Review the ECG signal
quality and correct if
necessary. Reposition the
ISO and J points.
ECG EQUIP MALF Hard Failure of the ECG
equipment or failure to
calibrate ECG.
Remove leadset.
Remove and re-insert
batteries. Let Self-
Test complete before
reinserting leadset.
Replace the
transceiver.
Contact Service.
ECG LEADS OFF
Note—This INOP may also
be configured to display as
a Red or Yellow Technical
Alarm.
Hard Multiple leads are off. Reattach ECG leads to
patient.
<electrode> LEAD OFF Hard Single lead is off.
Note—If primary lead is
MCL, lead will be
identified as V/C in INOP
text.
Reattach ECG lead to
patient.
INVALID LEADSET Hard Wrong leadset.
Leadset inserted
incorrectly.
Bad lead selection
switches in the
transceiver.
Check for appropriate
leadset.
Check for correct
insertion.
Contact Service.
NO ARRHYTHMIA Soft All basic ST/AR ECG
alarms have been turned
off.
Turn arrhythmia alarms on.
Technical Alarms (Inoperative Conditions)
Alarm Text Priority Condition What to do