Positioning ECG Electrodes
6-20 ECG Monitoring
V7 on the left posterior axillary line, straight line
from V6
V8 on the left midscapular line, straight line from
V7
V9 on the left paraspinal line, straight line from V8
V3R midway between the V1 and V4R electrode
positions
V4R on the fifth intercostal space at the right
midclavicular line
V5R on the right anterior axillary line, straight line
from V4R