User's Manual
Table Of Contents
- Introducing IntelliVue Telemetry
- Product Safety
- Transceiver Controls
- Basic Operation
- Alarms
- ECG Monitoring
- ST/AR Arrhythmia Monitoring
- SpO2 Monitoring
- Telemetry Functions at the Information Center & TeleMon
- Telemetry Functions at the Information Center
- Telemetry Controls in the Patient Window
- Locating the Transceiver (Find Device)
- Viewing Device Location in the Patient Window (optional)
- Viewing Device Location History (optional)
- Using the Device Location Client (optional)
- Patient Configurable Settings in Telemetry Setup
- Unit- Configurable Settings
- RF Auto Shutoff
- Transceiver Operation when Connected to TeleMon
- Telemetry Functions at the Information Center
- Pairing Monitoring Devices
- Maintenance, Cleaning & Troubleshooting
- Safety Standards & Specifications
- Accessories
- Sales and Support Offices
ST/AR Arrhythmia Algorithm
7-2 ST/AR Arrhythmia Monitoring
ST/AR Arrhythmia Algorithm
Indications
for Use
The ST/AR Arrhythmia Algorithm is indicated for use in instances where the
clinician decides to monitor cardiac arrhythmias of adult and pediatric patients
and/or the ST segment of adult patients to gain information for treatment,
monitor the adequacy of treatment, or to exclude causes of symptoms.
Safety
Information
WarningWarning
During complete heart block or pacemaker failure (to pace or capture), tall
P-waves (greater than 1/5 of the average R-wave height) can be erroneously
counted by the arrhythmia algorithm, resulting in missed detection of
cardiac arrest.
WarningWarning
Learning
If you initiate learning during ventricular rhythm, the ectopics can be
incorrectly learned as the normal QRS complex. This can result in missed
detection of subsequent events of V-Tach and V-Fib.