User's Manual
Table Of Contents
- Introducing IntelliVue Telemetry
- Product Safety
- Transceiver Controls
- Basic Operation
- Alarms
- ECG Monitoring
- ST/AR Arrhythmia Monitoring
- SpO2 Monitoring
- Telemetry Functions at the Information Center & TeleMon
- Telemetry Functions at the Information Center
- Telemetry Controls in the Patient Window
- Locating the Transceiver (Find Device)
- Viewing Device Location in the Patient Window (optional)
- Viewing Device Location History (optional)
- Using the Device Location Client (optional)
- Patient Configurable Settings in Telemetry Setup
- Unit- Configurable Settings
- RF Auto Shutoff
- Transceiver Operation when Connected to TeleMon
- Telemetry Functions at the Information Center
- Pairing Monitoring Devices
- Maintenance, Cleaning & Troubleshooting
- Safety Standards & Specifications
- Accessories
- Sales and Support Offices
How the QT Analysis Algorithm Works
7-20 ST/AR Arrhythmia Monitoring
Select the QT Lead
The QT Lead field allows you to select which
leads to analyze when calculating the QT
parameters. To select the desired lead by
clicking on the QT Lead drop down arrow then
highlighting the lead from the list that displays.
Choose:
• All if you want a global QT measurement
based on all available leads. For standard
placement leads I, II, III, V and V1 through
V6 are used. For EASI placement directly
acquired AI, AS, and ES leads are used.
Note—This may include leads that are not being
stored.
• Primary if you want to use the primary
lead for the QT measurement. If the primary
lead becomes unavailable or is changed the
QT measurement continues with the new
primary lead.
• A single lead from the list to use that lead
for QT measurement. If the lead you select
becomes unavailable QT monitoring stops.
Note—The V7, V8, V9, V3R, V4R or V5R
leads are not available for single lead
selection. These leads are processed,
however, when you select Primary in the
QT Lead field.
Adjustment Description