How the QT Analysis Algorithm Works
ST/AR Arrhythmia Monitoring
7-21
Set the QT Baseline
To quantify changes in the QTc value you can
set a QTc baseline. Select the Set QT Baseline
button to replace the baseline QTc value with the
current QTc value. If a baseline has not been set
the Information Center sets the baseline to the
first valid value after measuring QT for five
minutes. Setting a new baseline discards the
previous baseline.
Notes—
Since the dQTc alarm is based on the
difference between the baseline and the
current value, setting an inappropriate new
baseline may prevent a dQTc alarm from
being generated.
Turning QT analysis off does not clear the
baseline value. This allows you to turn QT
analysis off during prolonged arrhythmias,
such as bigeminy, without losing the
baseline.
Discharging a patient clears the baseline.