User's Manual

Table Of Contents
Telemetry Functions at the Information Center
9-8 Telemetry Functions at the Information Center & TeleMon
Suppress
SpO
2
INOPs
with NBP
Enable/disable the SpO
2
algorithm to suppress sending
technical alarms from the
transceiver during an NBP
measurement for 60 seconds.
WarningWarning
If you measure SpO
2
on a limb
that has an inflated NBP cuff, a
non-pulsatile SpO
2
technical
alarm can occur. If the monitor
is configured to suppress this
alarm, there can be a delay of
up to 60 seconds in indicating
critical patient status, such as
sudden pulse loss or hypoxia.
enable
disable
enable
Pleth Wave Enable/disable the transmission
of the Pleth wave (and its
subsequent display) to the
Information Center. For
Continuous SpO
2
mode only.
Note—When enabled, the Pleth
wave replaces the Vb wave in the
Patient Window during 6-lead
monitoring.
enable
disable
disable
(Pleth is not
displayed.)
Pulse Enable/disable display of the
Pulse rate at the Information
Center. For Continuous SpO
2
mode only.
enable
disable
disable
(Pulse rate is
not
displayed.)
Patient-Configurable Settings in Telemetry Setup
Control Function Setting Choices
Factory
Default