Networked Devices
Pairing Monitoring Devices
10-15
Paired Device
Synchronized
Alarm
Settings
(Networked)
If the patient’s ECG is initially being measured with a patient monitor, and
then the patient is connected to the transceiver for monitoring, the Information
Center will use the patient monitor settings for the transceiver. When the
initial ECG source is the transceiver, and then the patient is connected to the
monitor, the Information Center uses its Telemetry Setup settings. The
following settings will be synchronized:
Heart Rate HR/Pulse Alarm On/Off, Heart Rate High/Low Limit
ECG Primary Lead, Secondary Lead, Va Lead, Vb Lead
Arrhythmia Analysis Mode, Asystole Threshold, Pause
Threshold, VTach HR, VTach Run, PVCs/min, Vent.
Rhythm, SVT HR, SVT Run, PVCs/min On/Off,
Pacer not Capture On/Off, Pacer not Pace On/Off,
Non-sustain On/Off, Vent. Rhythm On/Off, Run
PVCs On/Off, Pair PVCs On/Off, Missed Beat On/
Off, Pause On/Off, R-on-T On/Off, Vent. Bigeminy
On/Off, Vent. Trigeminy On/Off, Multiform PVCs
On/Off, Irregular HR On/Off, SVT On/Off, Afib On/
Off
ST ST Analysis On/Off, ST Alarm On/Off, ISO point, J
point, ST point, ST Priority List, Single ST Alarm
Limit, Multi ST Alarm Limit
QT QT Analysis On/Off, QTc High On/Off, QTc High
Alarm Limit, dQTc High On/Off, dQTc High Alarm
Limit, QT Lead, QTc Correction Formula, QT
Baseline
SpO
2
T SpO
2
Alarms On/Off, SpO
2
Alarm Limits
NBP Alarm Suppression On/Off, Pulse (SpO
2
) On/
Off