General Safety
Product Safety
2-5
WarningWarning
Short-range radio connections are subject to interruption due to
interference from other radio sources in the vicinity, including microwaves,
bluetooth devices, and DECT phones. Outside the frequency band and 5%
above and below, i.e. the exclusion band according to IEC 60601-1-2, the
short-range radio connection is immune up to 3V/m in the frequency range
from 80MHz to 2.0 GHz and up to 1V/m in the frequency range from 2.0 to
2.3 GHz. Depending on the strength and duration of the interference, the
interruption may occur for an extended period. Any interruption of the
signal due to interference, moving out of range, or for other reasons is
indicated with a Tele Disconnected INOP message.